Model | Specification |
Internal Dim. W x H x D (cm) | 50x60x50 |
External Dim. W x H x D (cm) | 95x150x105 |
* Temp. & Humi. Range | -40〫C~ +150〫C & 10RH% ~98% RH |
Accelerated aging of sterile barrier systems for medical devices(Standard Guide: ASTM F1980-16) | |||
Conversion table for temperature vs testing time | |||
Temp( 〫C) | Accelerated aging time (Week) |
Equivalent real time aging (Year) |
(1)Ambient temperature:25〫C (2)Q10=2.(Conservative aging factor) |
55〫C | 6.5 Week | 1 Year | |
55〫C | 19.5 Week | 3 Year | |
55〫C | 32.5 Week | 5 Year | |
*Temperature range 50,55或60〫C |
Environmental test (Standard Guide: IEC 60068-2. GB/T 2423) | ||
Items | standard | Object |
Cold test | IEC 60068-2-1 GB/T 2423.1 |
Testing for electric and electronic products operation capacity under cold environment. |
Dry heat test | IEC 60068-2-2 GB/T2423.2 |
Testing for electric and electronic products operation capacity under heat environment. |
Damp heat, steady state test | IEC 60068-2-78 GB/T 2423.3 |
Testing for electric and electronic products operation capacity under damp heat and steady state environment. |
Composite temperature / Humidity cyclic test |
IEC 60068-2-38 GB/T 2423.34 |
Testing for electric and electronic products operation capacity under composite temperature /humidity cyclic environment. |
Storage product temperature cycle test and temperature range: | ||